Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms

August 25, 2026
in Medicine
Ophelia Keating
By Ophelia Keating Scienmag Editorial Profile - Health Services Research
Reading Time: 2 mins read
0
Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms

Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms

65
SHARES
587
VIEWS
Share on FacebookShare on Twitter
ADVERTISEMENT

Study summary

This retrospective single-center study evaluated plasma neurofilament light chain (pNfL) measured in 253 people with multiple sclerosis using two automated assays:

  • Roche Elecsys® NfL on cobas e801
  • Fujirebio Lumipulse® G NfL Blood on LUMIPULSE G600

Main findings

  • The assays were strongly correlated on the raw scale:
    • Pearson correlation: r = 0.879
  • However, they showed poor absolute agreement:
    • Concordance correlation coefficient: CCC = 0.107
    • Mean concentrations differed substantially:
      • Lumipulse: 14.02 pg/mL
      • Elecsys: 1.50 pg/mL
  • Therefore, the two assays should not be treated as interchangeable using raw pg/mL values.

Proposed conversion

The authors derived a directional Passing–Bablok conversion:

[
\text{Lumipulse-equivalent pNfL}

1.97 + 11.171 \times \text{Elecsys pNfL}
]

This equation is intended for conversion from Elecsys to Lumipulse values, within the analytical range studied. It should not automatically be assumed to work in the reverse direction or outside that range.

Simoa-referenced harmonization

Published equations were used to convert both platforms to Simoa-equivalent values:

  • Elecsys-derived Simoa value:

[
6.31 \times \text{Elecsys} + 2.33
]

  • Lumipulse-derived Simoa value:

[
0.94 \times \text{Lumipulse} + 1.30
]

Alignment improved after conversion, with a regression coefficient of approximately 1.03, suggesting better comparability on the harmonized scale. However, these conversions are literature-based and do not establish that either assay is clinically identical to Simoa.

Clinical validation of Elecsys

Higher Elecsys pNfL was associated with greater disability:

  • EDSS coefficient: 0.302
  • BMI-adjusted model: 0.335

Elecsys pNfL was also lower among treated than untreated patients, supporting expected biological and clinical validity in routine MS care.

Clinical interpretation

The practical message is:

  1. Do not compare raw Elecsys and Lumipulse pNfL concentrations directly.
  2. A patient switching platforms may appear to have a large change in pNfL solely because of assay bias.
  3. During platform transitions, clinicians should:
    • Continue interpreting results using the assay-specific reference range;
    • Prefer longitudinal monitoring on the same platform;
    • Consider conversion equations cautiously when necessary;
    • Avoid interpreting converted values as fully interchangeable measurements.

Important limitations

  • Retrospective, single-center design.
  • All samples were measured in singlicate.
  • Precision estimates such as cumulative intra- and inter-session CVs were not calculated.
  • Only one reagent lot per platform was used.
  • The Simoa conversions were based on published formulas rather than a direct three-platform calibration in the same samples.
  • The Elecsys-to-Lumipulse equation is directional and may not generalize beyond the observed concentration range.
  • Treatment status and disease characteristics may confound clinical associations.
  • Improved regression alignment after conversion does not necessarily demonstrate complete clinical agreement or equivalent decision thresholds.

Overall, the study supports clinical validity of Elecsys pNfL in MS, but emphasizes that Elecsys and Lumipulse values are assay-specific and cannot be directly substituted on the raw pg/mL scale.

Subject of Research: Medicine

Article Title: Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms

Article References: Criscuolo, V., Nicolella, V., Polito, C., Sirica, R., Fiorenza, M., Sorvillo, D., Cave, I. D., Carotenuto, A., Petracca, M., Lanzillo, R., Castaldo, G., Morra, V. B., Terracciano, D., & Moccia, M. (2026). Real-world Roche Elecsys® and Fujirebio Lumipulse® inter-platform comparability and clinical validation of plasma neurofilament light chain in multiple sclerosis. Journal of Neurology, 273(9), Article 546. https://doi.org/10.1007/s00415-026-14068-6

Image Credits: AI Generated

DOI: 10.1007/s00415-026-14068-6

Keywords: analytical range and assay interchangeability, assay conversion equations, assay correlation and agreement, biomarker standardization in multiple sclerosis, clinical validation of plasma NfL, Fujirebio Lumipulse NfL assay, neurodegeneration biomarker analysis, plasma neurofilament light chain testing, pNfL measurement comparison, retrospective single-center biomarker study, Roche Elecsys NfL assay, Simoa platform harmonization

Cite Scienmag News

Ophelia Keating. (August 25, 2026). Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms. Scienmag. https://scienmag.com/real-world-study-validates-comparable-plasma-nfl-testing-across-roche-and-fujirebio-platforms/

Ophelia Keating. "Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms." Scienmag, 25 August 2026, https://scienmag.com/real-world-study-validates-comparable-plasma-nfl-testing-across-roche-and-fujirebio-platforms/. Accessed 3 September 2026.

Ophelia Keating. "Real-World Study Validates Comparable Plasma NfL Testing Across Roche and Fujirebio Platforms." Scienmag. August 25, 2026. https://scienmag.com/real-world-study-validates-comparable-plasma-nfl-testing-across-roche-and-fujirebio-platforms/

Tags: analytical range and assay interchangeabilityassay conversion equationsassay correlation and agreementbiomarker standardization in multiple sclerosisclinical validation of plasma NfLFujirebio Lumipulse NfL assayneurodegeneration biomarker analysisplasma neurofilament light chain testingpNfL measurement comparisonretrospective single-center biomarker studyRoche Elecsys NfL assaySimoa platform harmonization
Share26Tweet16
Previous Post

Early Treatment Attrition After Recurrence Following Neoadjuvant Chemo-Immunotherapy in Triple-Negative Breast Cancer

Next Post

Mediterranean Diet and Pancreatic Cancer Risk Studied in 2.3 Million Participants

Related Posts

Restless Legs in Pregnancy Triples Risk of Perinatal Depression, Study Finds
Medicine

Restless Legs in Pregnancy Triples Risk of Perinatal Depression, Study Finds

September 3, 2026
Occupational Therapists as Social Prescribers: Insights from Swedish Primary Care
Medicine

Occupational Therapists as Social Prescribers: Insights from Swedish Primary Care

September 3, 2026
When Clinical AI Outpaces Human Oversight
Medicine

When Clinical AI Outpaces Human Oversight

September 3, 2026
Workplace stress and healthy aging: new insights from the Semmelweis study
Medicine

Workplace stress and healthy aging: new insights from the Semmelweis study

September 3, 2026
Worldwide study shows limits to how long humans can live
Medicine

Worldwide study shows limits to how long humans can live

September 3, 2026
Endoscopic robot with deep learning path planning for liver biopsy
Medicine

Endoscopic robot with deep learning path planning for liver biopsy

September 3, 2026
Next Post
Mediterranean Diet and Pancreatic Cancer Risk Studied in 2.3 Million Participants

Mediterranean Diet and Pancreatic Cancer Risk Studied in 2.3 Million Participants

  • Mothers who receive childcare support from maternal grandparents show more optimized

    Mothers who receive childcare support from maternal grandparents show more parental warmth, finds NTU Singapore study

    27656 shares
    Share 11059 Tweet 6912
  • University of Seville Breaks 120-Year-Old Mystery, Revises a Key Einstein Concept

    1061 shares
    Share 424 Tweet 265
  • Bee body mass, pathogens and local climate influence heat tolerance

    682 shares
    Share 273 Tweet 171
  • Researchers record first-ever images and data of a shark experiencing a boat strike

    546 shares
    Share 218 Tweet 137
  • Groundbreaking Clinical Trial Reveals Lubiprostone Enhances Kidney Function

    531 shares
    Share 212 Tweet 133
Science

Embark on a thrilling journey of discovery with Scienmag.com—your ultimate source for cutting-edge breakthroughs. Immerse yourself in a world where curiosity knows no limits and tomorrow’s possibilities become today’s reality!

RECENT NEWS

  • China’s CACA Guidelines Redefine Cancer Care With Holistic Integrative Assessment
  • Advances in detecting and regulating Alternaria toxins in tomato products
  • Tests May Underestimate Cognitive Skills of Deaf Young Adults
  • Yeast strains differ in dough gas cell stability and bread crumb structure

Categories

  • Agriculture
  • Anthropology
  • Archaeology
  • Athmospheric
  • Biology
  • Biotechnology
  • Blog
  • Bussines
  • Cancer
  • Chemistry
  • Climate
  • Earth Science
  • Editorial Policy
  • Marine
  • Mathematics
  • Medicine
  • Pediatry
  • Policy
  • Psychology & Psychiatry
  • Science Education
  • Social Science
  • Space
  • Technology and Engineering

Subscribe to Blog via Email

Enter your email address to subscribe to this blog and receive notifications of new posts by email.

Join 5,151 other subscribers

© 2025 Scienmag - Science Magazine

Welcome Back!

Login to your account below

Forgotten Password?

Retrieve your password

Please enter your username or email address to reset your password.

Log In
No Result
View All Result
  • HOME
  • SCIENCE NEWS
  • CONTACT US

© 2025 Scienmag - Science Magazine

Discover more from Science

Subscribe now to keep reading and get access to the full archive.

Continue reading