NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test

November 10, 2026
in Medicine
Ophelia Keating
By Ophelia Keating Scienmag Editorial Profile - Health Services Research
Reading Time: 3 mins read
0
NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test

NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test

65
SHARES
587
VIEWS
Share on FacebookShare on Twitter
ADVERTISEMENT

In a controlled bench-top evaluation using a two-dimensional phantom, researchers observed no visually detectable pointer deviation on the display of the NextAR surgical navigation system across tests utilizing C-arm, O-arm, and computed tomography imaging. The investigation, detailed in BMC Musculoskeletal Disorders by Richard Zaccaria and colleagues, evaluated whether an operator could identify an offset between the navigated pointer tip and fixed target points on the system’s three-dimensional display. The primary finding demonstrated an overall deviation rate of zero across thirty total registration trials, though the researchers emphasized that the endpoint was strictly visual rather than an independently measured metric.

The accuracy of surgical navigation systems remains critical for their safe and effective use during clinical procedures. For this investigation, the research team examined whether a positional deviation of the pointer tip from a one-millimetre target could be visually identified on the display of the NextAR navigation platform, manufactured by Medacta International SA. The experimental design specifically explored whether visual alignment varied when image-to-patient registration was performed using different imaging modalities, and whether any detectable error changed in relation to the distance between a fixed fiducial reference marker and the evaluated target.

To evaluate pointer alignment under controlled conditions, the researchers constructed a flat two-dimensional phantom consisting of millimetre graph paper mounted on a wooden board. The test grid featured two parallel rows containing twenty steel nails each, with every nail possessing a one-millimetre head. Adjacent nails along each row were positioned ten millimetres apart, while the two parallel rows were separated by a distance of one hundred and twenty millimetres. A fiducial reference marker was anchored midway between the two rows at the starting edge of the grid to define the coordinate origin for all subsequent measurements.

The experimental phantom was imaged across three distinct radiographic modalities: a Ziehm Vision RFD 3D C-arm, a Medtronic O-arm MVS O2, and a Siemens SOMATOM Drive computed tomography scanner. For each imaging modality, the team conducted ten separate image acquisitions, with each acquisition immediately followed by a new image-to-patient registration sequence. Within the reconstructed volume of each scan, the operator positioned the physical tip of the navigated pointer onto each nail head and visually inspected the NextAR three-dimensional display interface to determine whether any positional discrepancy appeared.

The primary study endpoint, defined as visually detectable pointer deviation, was characterized as a visible offset of the pointer tip from the centre of the nail head on the navigation screen. Because the NextAR system does not output a numerical deviation value during this process, the evaluation relied entirely on visual monitoring rather than an independently measured target registration error. No artificial registration error was introduced into the system, and the physical pointer tip was not tracked using external reference measurement instruments. The authors noted that the visual detection threshold was approximately one millimetre, constrained by the dimensions of the nail heads and the millimetre paper grid, and was neither calibrated nor verified independently.

Across all thirty conducted acquisitions, the operator detected zero instances of visible pointer offset on the system display. The observed rate of detectable deviation was zero out of ten acquisitions for each modality individually, corresponding to an exact two-sided Clopper–Pearson ninety-five percent confidence interval of 0 to 30.9 percent. Across the pooled thirty acquisitions, the overall rate of zero out of thirty yielded an exact ninety-five percent confidence interval of 0 to 11.6 percent under the fixed bench-top setup. The maximum fiducial-to-target distances assessed reached 143 millimetres with the C-arm, 152 millimetres with the O-arm, and 199 millimetres with computed tomography.

Despite the absence of visible deviations, the investigators noted essential constraints inherent to the bench-top setup. Because the evaluation relied exclusively on a visual observation threshold limited to approximately one millimetre without external reference validation, the findings do not establish submillimetre accuracy, demonstrate equivalence among the three imaging systems, confirm the absence of distance-related error effects, or guarantee clinical reliability. The authors concluded that thorough verification in anatomically representative cadaveric models and clinical surgical environments remains necessary before drawing conclusions regarding patient applications.

Subject of Research: Visual accuracy evaluation of the NextAR surgical navigation system

Article Title: Visually detectable pointer deviation of the nextar surgical navigation system with c-arm, o-arm and ct imaging: a bench-top study using a two-dimensional phantom

Article References: Zaccaria, R., Suero, E., Lampert, C., Simon, D., Neudeck, R., Hauck, S., Weigert, A., Reidler, P., Böcker, W., & Wegener, B. (2026). Visually detectable pointer deviation of the nextar surgical navigation system with c-arm, o-arm and ct imaging: a bench-top study using a two-dimensional phantom. BMC Musculoskeletal Disorders. https://doi.org/10.1186/s12891-026-10696-1

Image Credits: AI Generated

DOI: 10.1186/s12891-026-10696-1

Keywords: surgical navigation, NextAR, bench-top study, two-dimensional phantom, imaging modalities, accuracy assessment, computer-assisted surgery, fiducial registration, Visually, detectable, pointer, deviation

Cite Scienmag News

Ophelia Keating. (November 10, 2026). NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test. Scienmag. https://scienmag.com/nextar-surgical-navigation-shows-no-visually-detectable-pointer-deviation-in-bench-test/

Ophelia Keating. "NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test." Scienmag, 10 November 2026, https://scienmag.com/nextar-surgical-navigation-shows-no-visually-detectable-pointer-deviation-in-bench-test/. Accessed 10 November 2026.

Ophelia Keating. "NextAR Surgical Navigation Shows No Visually Detectable Pointer Deviation in Bench Test." Scienmag. November 10, 2026. https://scienmag.com/nextar-surgical-navigation-shows-no-visually-detectable-pointer-deviation-in-bench-test/

Tags: accuracy assessmentbench-top studycomputer-assisted surgerydetectableDeviationfiducial registrationimaging modalitiesNextARpointersurgical navigationtwo-dimensional phantomvisually
Share26Tweet16
Previous Post

Systematic Review Finds Limited Evidence for Dementia Apps

Next Post

Chromosome-Level Genome Assembly Unveils Genetic Architecture of Legume Pod Borer Pest

Related Posts

PGK1 Expression Associates With Overall Survival in TCGA Breast Cancer Cohort
Cancer

PGK1 Expression Associates With Overall Survival in TCGA Breast Cancer Cohort

November 10, 2026
High DIPSS Risk Linked to Reduced Stem Cell Transplant Survival in Myelofibrosis
Medicine

High DIPSS Risk Linked to Reduced Stem Cell Transplant Survival in Myelofibrosis

November 10, 2026
Lanthanum-135 SSTR2 Radioligands Show No In Vitro Auger Therapy Efficacy
Medicine

Lanthanum-135 SSTR2 Radioligands Show No In Vitro Auger Therapy Efficacy

November 10, 2026
Vascular Injury Extent Predicts Early Digit Survival After Replantation
Medicine

Vascular Injury Extent Predicts Early Digit Survival After Replantation

November 10, 2026
Nasopharyngeal Swabs Show High Specificity for West Nile Virus Detection
Medicine

Nasopharyngeal Swabs Show High Specificity for West Nile Virus Detection

November 10, 2026
Vessel Diameter and Complex Anatomy Predict Pulmonary Malformation Reopening After Embolisation
Medicine

Vessel Diameter and Complex Anatomy Predict Pulmonary Malformation Reopening After Embolisation

November 10, 2026
Next Post
Chromosome-Level Genome Assembly Unveils Genetic Architecture of Legume Pod Borer Pest

Chromosome-Level Genome Assembly Unveils Genetic Architecture of Legume Pod Borer Pest

  • Mothers who receive childcare support from maternal grandparents show more optimized

    Mothers who receive childcare support from maternal grandparents show more parental warmth, finds NTU Singapore study

    27656 shares
    Share 11059 Tweet 6912
  • University of Seville Breaks 120-Year-Old Mystery, Revises a Key Einstein Concept

    1061 shares
    Share 424 Tweet 265
  • Bee body mass, pathogens and local climate influence heat tolerance

    682 shares
    Share 273 Tweet 171
  • Researchers record first-ever images and data of a shark experiencing a boat strike

    546 shares
    Share 218 Tweet 137
  • Groundbreaking Clinical Trial Reveals Lubiprostone Enhances Kidney Function

    531 shares
    Share 212 Tweet 133
Science

Embark on a thrilling journey of discovery with Scienmag.com—your ultimate source for cutting-edge breakthroughs. Immerse yourself in a world where curiosity knows no limits and tomorrow’s possibilities become today’s reality!

RECENT NEWS

  • Transformer Model Achieves 96.28% Accuracy in Tennis Stroke Recognition
  • PGK1 Expression Associates With Overall Survival in TCGA Breast Cancer Cohort
  • High DIPSS Risk Linked to Reduced Stem Cell Transplant Survival in Myelofibrosis
  • Lanthanum-135 SSTR2 Radioligands Show No In Vitro Auger Therapy Efficacy

Categories

  • Agriculture
  • Anthropology
  • Archaeology
  • Athmospheric
  • Biology
  • Biotechnology
  • Blog
  • Bussines
  • Cancer
  • Chemistry
  • Climate
  • Earth Science
  • Editorial Policy
  • Marine
  • Mathematics
  • Medicine
  • Pediatry
  • Policy
  • Psychology & Psychiatry
  • Science Education
  • Science News
  • Social Science
  • Space
  • Technology and Engineering

Subscribe to Blog via Email

Enter your email address to subscribe to this blog and receive notifications of new posts by email.

Join 5,150 other subscribers

© 2026 Scienmag - Science Magazine

Welcome Back!

Login to your account below

Forgotten Password?

Retrieve your password

Please enter your username or email address to reset your password.

Log In
No Result
View All Result
  • HOME
  • SCIENCE NEWS
  • CONTACT US

© 2026 Scienmag - Science Magazine

Discover more from Science

Subscribe now to keep reading and get access to the full archive.

Continue reading