Robotic surgical platforms generate continuous streams of digital procedural data that offer opportunities to assess surgical performance, according to a retrospective study evaluating rectal transection techniques in robotic rectal cancer surgery. Researchers investigated whether objective log data from the da Vinci SureForm stapling system could characterize how surgeons manipulate and deploy surgical staplers during anterior resection procedures. Although robotic platforms record instrument movements automatically, quantitative evaluations of robotic stapler utilization during rectal transections have historically remained limited in surgical literature.
The clinical investigation, conducted by Yoshiko Bamba, Shimpei Ogawa, and Shigeki Yamaguchi, reviewed procedural logs from thirty-two patients who underwent robotic anterior resection for rectal cancer between 2022 and 2024. During these operations, surgeons utilized the da Vinci SureForm stapling system to transect the rectum. To evaluate instrument deployment, the researchers extracted detailed log metrics recorded by the system, focusing specifically on mechanical firing duration along with spatial orientation measures, namely pitch angle and yaw angle.
To analyze procedural variations, the investigators separated the patient cohort into two distinct clinical categories: a single-firing group and a multiple-firing group. This classification depended directly on the total number of stapler firings necessary to achieve complete rectal transection during surgery. Across the entire surgical series of thirty-two patients, the study evaluated a total of fifty-five individual stapler firings, capturing the mechanical parameters and instrument positioning associated with each transection attempt.
The extracted log data revealed consistent patterns in surgical instrument orientation across the evaluated operations. Across all fifty-five firings, the median pitch angle measured 46.4 degrees, with an interquartile range of 35.7 to 52.8 degrees and an overall range from 1.2 to 59.9 degrees. The authors noted that the majority of stapler firings fell within a pitch orientation of 40 to 55 degrees. Instrument yaw angles exhibited a median of 15.8 degrees, spanning an interquartile range of 9.5 to 25.3 degrees and an overall range between 0 and 59.9 degrees, with most firings concentrated within 0 to 20 degrees.
When comparing procedural outcomes, the researchers determined that patients requiring multiple firings experienced significantly longer total operative times compared to those undergoing transection with a single firing. Furthermore, multiple stapler firings tended to occur more frequently among patients diagnosed with lower rectal tumors and in cases where surgeons created diverting stomas. Despite these operational differences, statistical analyses demonstrated no significant variations in either pitch angles or yaw angles between the single-firing and multiple-firing groups.
The authors also detailed an individual case within the cohort that required four stapler firings to complete the transection. This complex case involved a patient presenting with locally advanced rectosigmoid cancer exhibiting bladder invasion following total neoadjuvant therapy.
In their conclusions, the investigators emphasized that SureForm stapler log data successfully enabled the objective characterization of firing duration and stapler orientation during robotic rectal transection. The authors identified these exploratory measurements as preliminary reference data for robotic stapler utilization in clinical practice. Because the findings derive from a single-center retrospective analysis, the researchers concluded that larger multicenter studies will be necessary to establish the broader clinical relevance of robotic stapler log metrics.
Subject of Research: Objective quantification of robotic stapler orientation and firing during rectal transection
Article Title: Objective quantification of rectal transection technique using SureForm™ stapler log data in robotic rectal cancer surgery
Article References: Bamba, Y., Ogawa, S., & Yamaguchi, S. (2026). Objective quantification of rectal transection technique using SureForm™ stapler log data in robotic rectal cancer surgery. Techniques in Coloproctology. https://doi.org/10.1007/s10151-026-03441-2
Image Credits: AI Generated
DOI: 10.1007/s10151-026-03441-2
Keywords: robotic rectal cancer surgery, rectal transection, SureForm stapler, stapler log data, robotic anterior resection, pitch angle, yaw angle, da Vinci surgical system, firing duration, multiple stapler firings, single stapler firing, lower rectal tumors, diverting stoma, operative time, total neoadjuvant therapy, rectosigmoid cancer, colorectal surgery, surgical technique quantification