A custom-built, digit-resolved grip dynamometer reproduces conventional handgrip strength outcomes with good agreement relative to a standard clinical Jamar dynamometer while successfully capturing finger-specific force measurements, according to an open-access study published in the Journal of NeuroEngineering and Rehabilitation. Developed by researchers Gregory M. Shaw, Andrew B. Dick, Thomas J. Boyle, Robert Williams, and Brian C. Clark, the novel measurement device was engineered to approximate standard Jamar testing conditions while providing the digit-level force resolution that is fundamentally unavailable from traditional single-output dynamometers.
The investigation focused on addressing the methodological limitations surrounding the measurement of the multi-finger force deficit, a biomechanical phenomenon defined as the reduction in force expressed during simultaneous multi-finger exertion relative to the summed force capacity of the individual digits. Although scientific research has linked the multi-finger force deficit to neural control mechanisms, previous studies have generally measured the deficit using experimental apparatuses and tasks that do not biomechanically match traditional handgrip strength testing configurations.
To resolve these biomechanical discrepancies, the research team designed and manufactured the custom digit-resolved grip dynamometer to quantify both whole-hand grip strength and individual digit-resolved forces within an architecture engineered to approximate standard Jamar testing conditions. By capturing individual finger forces during grip exertion, the device was specifically configured to generate the finger-specific force signals required to quantify multi-finger coordination features alongside standard whole-hand strength metrics in 22 human volunteers.
Before testing human participants, the researchers conducted known-mass dead-weight testing to assess the static load-measurement accuracy of the newly manufactured device independently of the clinical reference Jamar dynamometer. These dead-weight validation procedures revealed highly linear load measurement across the entire tested force range, with coefficient of determination values exceeding 0.999 for both devices. The bench evaluations demonstrated a mean relative error of positive 5.0 percent for the custom digit-resolved grip dynamometer and a mean relative error of negative 2.7 percent for the reference Jamar dynamometer.
In the volunteer testing protocol, the researchers established standardized force outcomes for both devices across repeated maximal whole-hand trials. For each dynamometer, peak handgrip strength was defined as the single highest force achieved across all maximal whole-hand trials, whereas mean handgrip strength was calculated from the final two maximal trials. The team then evaluated device association and agreement through Pearson correlation coefficients, intraclass correlation coefficients, and Bland–Altman analysis.
The comparative analyses demonstrated strong pooled correlations between the custom digit-resolved grip dynamometer and the reference Jamar dynamometer for both peak and mean handgrip strength measurements, with both metrics yielding a Pearson correlation coefficient of r = 0.99. Furthermore, peak handgrip strength displayed good statistical agreement across both testing conditions, recording an intraclass correlation coefficient of 0.872 with a bias of negative 0.56 percent in the dominant hand, alongside an intraclass correlation coefficient of 0.890 with a bias of negative 9.19 percent in the non-dominant hand.
Analyses of mean handgrip strength similarly demonstrated good agreement between the two dynamometers for both limbs. In the dominant hand, mean handgrip strength demonstrated an intraclass correlation coefficient of 0.805 and a mean bias of negative 3.74 percent, whereas the non-dominant hand exhibited an intraclass correlation coefficient of 0.861 and a median bias of negative 9.83 percent. The authors observed that the larger differences identified in the non-dominant hand are consistent with potential order, fatigue, and attentional effects arising from the study’s fixed testing sequence.
Based on these findings, the researchers concluded that the custom digit-resolved grip dynamometer serves as a viable measurement platform for conventional handgrip strength assessments and for quantifying multi-digit coordination features. While the observed differences between the instruments were generally modest and compatible with protocol order effects, the authors emphasized that device-related and protocol-related sources of variance cannot be fully separated in the present study design.
Subject of Research: Development and validation of a digit-resolved grip dynamometer for measuring multi-finger force coordination
Article Title: A digit-resolved grip dynamometer for quantifying multi-finger force coordination: development and validation against reference hand dynamometer
Article References: Shaw, G. M., B. Dick, A., Boyle, T. J., Williams, R., & Clark, B. C. (2026). A digit-resolved grip dynamometer for quantifying multi-finger force coordination: development and validation against reference hand dynamometer. Journal of NeuroEngineering and Rehabilitation. https://doi.org/10.1186/s12984-026-02204-2
Image Credits: AI Generated
DOI: 10.1186/s12984-026-02204-2
Keywords: digit-resolved grip dynamometer, handgrip strength, multi-finger force deficit, Jamar dynamometer, force coordination, dead-weight testing, intraclass correlation coefficients, Bland-Altman analysis, Pearson correlation, biomechanical testing, neural control mechanisms, dominant hand grip, non-dominant hand grip, static load measurement, peak handgrip strength, mean handgrip strength, single-output dynamometry
Cite Scienmag News
Ophelia Keating. (November 18, 2026). Digit-Resolved Grip Dynamometer Matches Clinical Reference Tool While Tracking Individual Fingers. Scienmag. https://scienmag.com/digit-resolved-grip-dynamometer-matches-clinical-reference-tool-while-tracking-individual-fingers/
Ophelia Keating. "Digit-Resolved Grip Dynamometer Matches Clinical Reference Tool While Tracking Individual Fingers." Scienmag, 18 November 2026, https://scienmag.com/digit-resolved-grip-dynamometer-matches-clinical-reference-tool-while-tracking-individual-fingers/. Accessed 18 November 2026.
Ophelia Keating. "Digit-Resolved Grip Dynamometer Matches Clinical Reference Tool While Tracking Individual Fingers." Scienmag. November 18, 2026. https://scienmag.com/digit-resolved-grip-dynamometer-matches-clinical-reference-tool-while-tracking-individual-fingers/