Magnetic component in e-cigarettes found to interfere with implantable cardioverter-defibrillator function

March 16, 2020
in Technology and Engineering
0
Share on FacebookShare on Twitter

Potential for unintended temporary programming and arrhythmic complications when an electronic cigarette is placed in close proximity to an implantable cardioverter-defibrillator or pacemaker reported in HeartRhythm Case Reports

Philadelphia, March 16, 2020 – An e-cigarette carried in the left breast shirt pocket of a patient with an implantable cardioverter-defibrillator (ICD) caused magnetic reversion, interrupting the ICD’s ability to detect and treat dangerous heart rhythm problems, clinicians report in HeartRhythm Case Reports, published by Elsevier. The patient was not aware that the e-cigarette has an integrated magnetic component, and it had suspended detection of heart rhythm problems by the ICD four times before he reported it to his healthcare team.

“To our knowledge this is the first reported case of magnetic reversion of an ICD by an e-cigarette,” stated senior investigator Usha B. Tedrow, MD, MPH, and lead authors Julie B. Shea, MS, RNCS, Martin Aguilar, MD, and William Sauer, MD, from the Cardiovascular Arrhythmia Service at Brigham and Women’s Hospital and Harvard Medical School, Boston, MA, USA. “Given the increasing use of e-cigarettes worldwide, recognition of this potentially serious interaction appears clinically important.”

A 48-year-old male with an ICD reported that he heard his device “beep” several times, both at home and at his office. There were no adverse symptoms associated with the beeping, and remote monitoring found the device was working normally. The patient denied any magnetic exposure, but data provided by the ICD manufacturer found four instances of magnet interactions with the device, corresponding with when the patient heard the tone.

Upon further questioning, the patient recalled using his e-cigarette (JUUL vape device), which he frequently stored in his left breast pocket, overlying the ICD. When the healthcare team held the device up to his ICD, it triggered the steady magnet tone. The patient was educated about the importance of keeping any type of magnet away from his ICD. JUUL’s website does recommend keeping e-cigarettes away from key cards, credit cards, and other items with magnetic strips, as well as pacemakers.

“Magnets are ubiquitous in commercially-available electronic devices. They can be integrated in ways that are difficult to recognize. Although manufacturers are not routinely required to specify the strength of the magnetic fields and safety information for interference with medical-grade devices, the general recommendation is that any portable electronic or magnetic device be kept at least six to 12 inches away from an implant,” noted Dr. Tedrow and Ms. Shea.

There are commercially available magnetic field meters, and even several smartphone applications, that can be used to estimate the strength of a magnet. Practically speaking, most cardiac implantable devices have a magnetic exposure upper limit of 10G, and manufacturers typically recommend a 2:1 safety margin for safe clinical operation. “As such, finding the distance at which the magnetic field is 5G or less would, in principle, provide adequate clearance for safe clinical operation of cardiac implantable devices,” Dr. Aguilar added.

“Practitioners should remain vigilant regarding the use of new technology by their ICD patients. In our case, there was no adverse effect from the interaction of the e-cigarette with the device, but if it had happened during a tachycardia episode, it could have had serious, perhaps even fatal consequences,” cautioned Dr. Tedrow and Ms. Shea.

###

Media Contact
Jane Grochowski
[email protected]
406-542-8397
http://dx.doi.org/10.1016/j.hrcr.2020.01.013

Tags: CardiologyElectrical Engineering/ElectronicsMedicine/HealthPublic HealthSmoking/Tobacco
Share27Tweet17Share5ShareSendShare
  • Body weight influences the chance of developing Polycystic Ovary Syndrome

    69 shares
    Share 28 Tweet 17
  • Protein supplement helps control Type 2 diabetes

    66 shares
    Share 26 Tweet 17
  • Study reveals potential target for treatment of diseases associated with mitochondrial DNA mutations

    65 shares
    Share 26 Tweet 16
  • New survey illustrates challenges associated with healthcare environmental hygiene in facilities worldwide

    67 shares
    Share 27 Tweet 17
  • Shared autonomous micro-mobility

    65 shares
    Share 26 Tweet 16
  • Discovery offers starting point for better gene-editing tools

    65 shares
    Share 26 Tweet 16
ADVERTISEMENT

About us

We bring you the latest science news from best research centers and universities around the world. Check our website.

Latest NEWS

Data contradict fears of COVID-19 vaccine effects on pregnancy and fertility

Long-duration energy storage beats the challenge of week-long wind-power lulls

‘Democracy’ governs mass jackdaw take-offs

Subscribe to Blog via Email

Enter your email address to subscribe to this blog and receive notifications of new posts by email.

Join 188 other subscribers

© 2022 Scienmag- Science Magazine: Latest Science News.

No Result
View All Result
  • HOME PAGE
  • BIOLOGY
  • CHEMISTRY AND PHYSICS
  • MEDICINE
    • Cancer
    • Infectious Emerging Diseases
  • SPACE
  • TECHNOLOGY
  • CONTACT US

© 2022 Scienmag- Science Magazine: Latest Science News.

Welcome Back!

Login to your account below

Forgotten Password?

Retrieve your password

Please enter your username or email address to reset your password.

Log In
Posting....